1 Overview

BR120-50-01 Supercritical CO2 Fluid Extraction Equipment is a (25+25+25+25)L/40Mpa four-extraction, two-divided CO2 circulation system for material extraction (solid or liquid) at high pressure and temperature, in a separator. The conditions were changed so that dissolved substances were resolved for the purpose of separation. The device is mainly composed of: extraction kettle, separation kettle, CO2 high pressure pump, entrainer pump, refrigeration system, CO2 storage tank, heat exchange system, purification system, flow meter, temperature control system, and safety protection device. Auxiliary equipment and systems can meet the requirements and flowchart requirements of the Super Supercritical CO2 Fluid Extraction main system.

2, the system composition

(1) Extraction system: including extraction kettle, heat exchanger, heating bath, temperature control, safety protection, pressure test, etc.

(2) Separation system: including separation kettle, heat exchanger, heating bath, temperature control, safety protection, pressure test, etc.

(3) Pressure system: including liquid CO2 pump, cold water circulation, safety protection, electrical contact protection, pressure test, etc.

(4) Regulator system: including filters, back pressure valves, etc.

(5) Liquid CO2 generation system: including refrigeration unit, liquid CO2 storage tank, gas-liquid separator, pressure test, etc.

(6) Entrainer system: including entrainer pump, entrainer addition tank, entrainer flow meter, etc.

(7) CO2 circulation system: including mixer, purifier, liquid CO2 flowmeter, etc.

(8) CO2 recovery system: Including CO2 recovery pump (optional for users), CO2 recovery pipeline, etc.

3, the system has the following major advantages

(1) The BR120-50-01 Supercritical CO2 Fluid Extraction Equipment  of the entire system is designed to facilitate the operation of the system. The appearance is beautiful, the noise is low, and the operation is convenient.

(2) The liquid CO2 booster pump is a three-piston reciprocating pump with a maximum output pressure of 40 MPa and a maximum flow rate of 1000 L/h. The pump head includes a cooling system and stepless frequency conversion.

(3) The heating method adopts the hot water circulation method, and the temperature control system adopts PID self-tuning, thereby improving the accuracy of temperature control and reducing the noise.

4, technical parameters

<table class="gridtable >

NO

Part Name

Parts Description

数量

1

Extraction kettle

With water jacketed circulation heating system, material: 0Cr18Ni9,

Adjustable temperature, with solid, liquid barrel, volume: 25L,

The highest working pressure: 50MPa, the highest temperature can be set: 85 °C,

4 set

2

Separator

With water jacketed circulation heating system, adjustable temperature,

Volume: 20L, material: 0Cr18Ni9,

Maximum working pressure: 30MPa, maximum settable temperature: 85°C

2 sets

3

Liquid CO2 high pressure pump

Form: three plunger, pump head with cooling system, maximum flow: 50L/h,

Variable frequency adjustable (with frequency converter), maximum working pressure: 50Mpa,

1 set

4

Entrainer pump

Form: three plunger, pump head with cooling system, adjustable frequency,

Maximum flow: 1000L/h, maximum working pressure: 40Mpa

1 set

5

Cooling System

Cooling capacity: 20000 Kcal/h, air-cooled,

Temperature control range: 2 °C ~ 8 °C, to meet the process requirements

1 set

6

CO2 storage tank

With water jacket cooling cooling system, material: 0Cr18Ni9,

Volume: 20L, pressure: 16MPa, including liquid level gauge

2 set

7

Heat Exchanger

Material: 0Cr18Ni9, Specification: Φ18×2 coil,

Maximum working pressure: 40MPa, water distribution jacket heating system

3 sets

8

filter

With inner liner desiccant, material: 0Cr18Ni9

Volume: 2L, pressure: 40MPa

1 set

9

Gas-liquid separator

Into the up and down the separation, material: 0Cr18Ni9,

Volume: 2L, pressure: 40MPa

1 set

10

mixer

ZR-I type, material: 0Cr18Ni9, volume: 2L, pressure: 40MPa

1 set

11

Liquid CO2 flowmeter

Form: Metal rotor flowmeter, measuring range: 0-1000L/h,

Pressure resistance: 10MPa, digital instantaneous and cumulative flow

1 set

12

Entrainer flowmeter

Form: Metal floater flowmeter, measuring range: 0-20L/h,

Pressure resistance: normal pressure, glass readings,

1 set

13

Heated water bath

Control range: adjustable from room temperature to 85°C (water bath)

Dynamic temperature control accuracy: ± 0.5 °C, digital display dual screen

6 sets

14

Regulator

Between the extraction and separation, control the extraction pressure,

Control pressure range: 0~40MPa

1 set

15

Precision pressure gauge

16MPa,40MPa,60MPa

7 sets

16

Electrode pressure meter

60MPa

1 set

17

Safety protection device

Three-fold protection of safety valves, smart meters, and electrode voltage meters

4 sets

18

Fittings, fittings, common pressure gauges, etc.

DW6, DW10, Pressure: 50MPa, Material: 0Cr18Ni9

1 set

19

Overall bracket

And control box

Spray treatment, configuration cylinder holder and cylinder heating jacket

1 set

20

Random Accessories

Seals, pipe valves, connectors, special tools, etc.

Batch 1